Yes, in the iT100/200 series. Jumper J10 exists inside the transmitter to select AC- or DC-coupling of the buffered sensor output. AC-coupling makes the output +/- with reference to common, DC-coupling makes the output +/- with respect to ½ of the power supply (not sensor BOV).

The iT300 series is automatically DC-coupled.